Specifications
|
Functions |
Ground resistance test, Loop resistance test |
|
Ambient Temperature and Humidity |
23℃±5℃,below 75%rh |
|
Power Supply |
DC 6V (4 AA alkaline dry batteries) |
|
Measure Method |
Mutual induction |
|
Resistance Resolution |
0.001Ω |
|
Current Resolution |
0.01mA |
|
Jaw Size |
55mm×32mm |
|
Ω + A synchronous display |
Ω + A synchronous display |
|
Ground Voltage Measure Function |
No |
|
Display Mode |
4-digit LCD display, black screen design |
|
LCD Size |
46mm×29mm |
|
Meter Size |
L/W/H:285mm×85mm×58mm |
|
Measure Time |
2times/second |
|
Data Storage |
300 groups,"MEM" storage indication, display "FULL" symbol indicates that storage is full |
|
Data Review |
"MR" symbol indicates when looking up data |
|
Overflow display |
"OL" symbol indicates over-range overflow |
|
Interference Test |
Automatically identify interference signals, the "NOISE" symbol indicates when the interference current is large |
|
Alarm Function |
Alarm when the measured value exceeds the alarm setting value |
|
Battery Voltage |
Real-time display of battery power, reminding timely charging when battery voltage is low |
|
Automatic shut-down |
Turn off after 5 minutes |
|
Power Consumption |
50mA Max |
|
Weight |
Meter:1180g (including battery) |
|
Working Temperature and Humidity |
-10℃~40℃;below 80%rh |
|
Storage Temperature and Humidity |
-20℃~60℃;below 70%rh |
|
Insulation Resistance |
Above 20MΩ (500V between circuit and case) |
|
Pressure resistance |
AC 3700V/rms(between circuit and case) |
|
External Magnetic Field |
<40A/m |
|
External Electric Field |
<1V/m |
|
Suitable for Safety Regulations |
IEC61010-1(CAT Ⅲ 300V、CAT IV 150V、Pollution degree2);IEC61010-031;IEC61557-1(earth resistance); |